| Operating temperature range | -20 ... 40°C |
| Temperature adjustment | digital |
| Temperature indication | digital |
| Internal temperature sensor | Pt100 |
| Resolution of display | 0.1 K |
| Temperature stability at -10°C | 0,2 K |
| Safety classification | Class I/NFL |
| Cooling power at 15°C at 0°C at -10°C at -20°C |
0.3 kW 0.2 kW 0.14 kW 0.07 kW |
| Refrigeration machine | air-cooled, natural refrigerant |
| Refrigerant | R290 |
| Refrigerant quantity | 0.04 kg |
| Circulation pump | Pressure- and suction pump |
| max. delivery | 20 l/min |
| max. delivery pressure | 0.2 bar |
| max. delivery (suction) | 17 l/min |
| Pump connection | G 1/4" female thread |
| min. filling capacity | 1.4 l |
| Volume of expansion | 2.6 l |
| Overall dimensions WxDxH ** | 225x360x380 mm |
| Net weight | 23 kg |
| Power supply requirement | 230V 1~ 50/60Hz |
| max. current | 2.8 A |
| Fuse (1 phase) | 4 A |
| min. ambient temperature | 5°C |
| max. ambient temperature | 40°C |
Water cooler, 115V
目录号
格式
价格
B02010003
1 unit
$5,100.00
This water cooling system allows continuous (Bioruptor® Standard or Plus) or regulated cooling (Bioruptor® Plus or Pico) of the Bioruptor sonication bath. The regulated valve (Single Cycle valve) ensures that water will only be replaced during the OFF cycle to avoid any interference between the water flow and the sonication process.